DIMENSIONS & SPECIFICATIONS
Technical Data
Technical Data
Ìý
Measuring methods | Potentriometric |
---|---|
Measuring cycle | at least 6 min. |
Measuring range |
NO3 - 0 - 1 to 0 - 500 mg N/l |
Precision | typ. <5% of measuring range (end of value) |
Drift | typ. <1% of measuring range (end of value) |
Reagent supply |
for approx. 3 weeks |
Number of measuring points | max. 6 |
Output signal |
|
Interfaces |
USB / Ethernet |
Relay contacts |
4/12 potential free contacts |
Digital inputs | 4/12 e.g. activating and deactivating of measuring points, system control |
Sample |
pressure-free Temperature:Ìý15 - 45°C (288 - 318 K) Flow:Ìý3 – 10 l/hÌýfree from suspendedÌýmatter and oil Connection:Ìýtube, flexibleÌý (ID 1.5 - 3 mm) |
Drain | pressure-freeÌýtube, flexibleÌý(ID 10 mm) |
Power supply |
85…264 VAC atÌý47…63 Hz |
Power consumption: |
~ 42 VA (max. 90 VA) |
Environmental temperature | 15 - 35°C (288 - 308 K) |
Installation | wall-mounted |
Protection classÌý(EN 60529) |
|
Weight |
housing with reagent cabinet 53 - 60 kg without reagents |
Dimensions (HEIGHT X WIDTH X DEPTH) |
housing: 700x600x320 mmÌý with reagent cabinet:Ìý1100x600x354 mm |